Command reference
R&S
®
ZNA
1214User Manual 1178.6462.02 ─ 29
CONTrol:GPIO<Port>:SENSe:CURRent? [<ALL>]
Returns the results of the current measurement on the related GPIO pin or ALL GPIO
pins.
Suffix:
<Port>
.
1, ..., 10
GPIO port number
If ALL currents are queried, this suffix is ignored.
Query parameters:
<ALL> ALL
Use ALL to measure the currents at all GPIO pins.
Return values:
<Results> Measured currents, either a single value or a comma-separated
list.
Default unit: A
Usage: Query only
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Voltage, Current" on page 665
CONTrol:GPIO:SENSe:SUMCurrent? <GpioPorts>
Returns the sum of the currents measured at the specified GPIO ports
Query parameters:
<GpioPorts> Comma-separated list of GPIO port numbers
The list must contain minimum two and maximum 10 numbers
between 1 and 10 in arbitrary order.
Example:
CONTrol:GPIO:SENSe:SUMCurrent? 1,2,3,4,5
returns the sum of the currents measured in GPIO ports 1 to 5
Usage: Query only
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Voltage, Current" on page 665
CONTrol<Ch>:GPIO:SENSe:TRIGger
Starts the voltage/current measurements on all GPIO pins.
The measurement time can be defined using CONTrol<Ch>:GPIO:TIME.
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted.
Usage: Event
Options: R&S ZN-B15/Z15 Var. 03
SCPI command reference