Command reference
R&S
®
ZNA
1217User Manual 1178.6462.02 ─ 29
CONTrol<Ch>:GPIO:TIME <MeasTime>
Sets the measurement time for the voltage/current measurements on the RFFE and
GPIO pins.
Same functionality as CONTrol<Ch>:RFFE:TEST:TIME.
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted: the measurement time
is valid for all channels.
Parameters:
<MeasTime> Measurement (= sampling = averaging) time
Range: 95 μs to 100 ms
*RST: 100 ms
Default unit: s
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Meas" on page 664
CONTrol<Ch>:GPIO<Port>:VOLTage[:DEFault] <Voltage>
Sets the output voltage of the respective GPIO port.
This voltage is applied using CONTrol<Ch>:GPIO:VOLTage:OUTPut.
Suffix:
<Ch>
.
Channel number
<Port> GPIO port number
Parameters:
<Voltage> Range: -7 to +15 V
Increment: 5 mV
Default unit: V
Manual operation: See "Voltage / Output Voltage" on page 663
CONTrol<Ch>:GPIO:VOLTage:OUTPut
Applies the current output voltage and current range (shunt) settings to the RFFE/
GPIO pins.
Same function as CONTrol<Ch>:RFFE:TEST:OUTPut on page 1222.
Suffix:
<Ch>
.
Channel number
Usage: Event
Manual operation: See "Apply" on page 663
SCPI command reference