Command reference
R&S
®
ZNA
1216User Manual 1178.6462.02 ─ 29
Example:
:CONTrol1:GPIO1:RANGe 20mA
:CONTrol1:GPIO1:SHUNt?
Returns 10 [Ω]
:CONTrol1:GPIO1:RANGe 2mA
:CONTrol:GPIO1:TEST:CLOCk:SHUNt?
Returns 100 [Ω]
:CONTrol1:GPIO1:RANGe 200uA
:CONTrol:GPIO1:SHUNt?
Returns 1000 [Ω]
:CONTrol1:GPIO9:RANGe 0
:CONTrol:GPIO9:SHUNt?
Returns 100000000 [Ω]
:CONTrol1:GPIO9:RANGe 100mA
:CONTrol:GPIO9:SHUNt?
Returns 1 [Ω]
Usage:
Query only
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Range / Shunt" on page 664
Table 7-11: Pins 1 to 8
Range
2 μA 20 μA 200 μA 2 mA 20 mA
Shunt 100 kΩ 10 kΩ 1 kΩ 100 Ω 10 Ω
Table 7-12: Pins 9 and 10
Range
0 mA 100 mA
Shunt 100 MΩ 1 Ω
CONTrol<Ch>:GPIO<Port>[:STATe] <EnableInSequence>
Enables/disables GPIO port <port> in the Sweep Sequencer for channel <Ch> (see
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:VOLTage and CONTrol<Ch>:
SEGMent<SegNr>:SEQuence<SeqNr>:GPIO<Port>:VOLTage).
Suffix:
<Ch>
.
Channel number
<Port> GPIO port number
Parameters:
<EnableInSequence> Enabled state
Manual operation: See "Seq." on page 663