Command reference
R&S
®
ZNA
1221User Manual 1178.6462.02 ─ 29
Options: R&S ZN-B15 Var. 03
Manual operation: See "Voltage, Current" on page 662
CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk:RANGe <Clock Current Range>
CONTrol<Ch>:RFFE<Bus>:TEST:DATA:RANGe <Data Current Range>
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:RANGe <CurrentRange>
Defines the (upper bound of the) current range for the voltage/current measurement on
the respective RFFE pin. The analyzer firmware automatically selects a suitable shunt
resistance, which can be queried using CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk|
DATA|VIO:SHUNt.
Suffix:
<Ch>
.
Channel number
<Bus> RFFE bus interface 1 or 2
Parameters:
<CurrentRange> Range:
2 x 10
n
μA with n=1,...,5
Default unit: A
Example: See CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk|DATA|
VIO:SHUNt
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Range / Shunt" on page 661
CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk:SHUNt?
CONTrol<Ch>:RFFE<Bus>:TEST:DATA:SHUNt?
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:SHUNt?
Returns the shunt resistance (in Ω) selected by the analyzer firmware for the config-
ured current range (see CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk|DATA|
VIO:RANGe).
The dependency between current range and shunt resistance is displayed in the table
below.
Suffix:
<Ch>
.
Channel number
<Bus> RFFE bus interface 1 or 2
SCPI command reference