Command reference
R&S
®
ZNA
1330User Manual 1178.6462.02 ─ 29
Manual operation: See "Cable & Splitter Loss" on page 907
[SENSe:]CONVerter<Port>:CSSLope <Factor>
Default slope factor for the local port of converter <Port>.
Suffix:
<Port>
.
Port number
Parameters:
<Factor> Slope factor
*RST: 0
Default unit: dB/GHz
Options: R&S ZNA-K8
Manual operation: See "Slope" on page 907
[SENSe:]CONVerter:DEFinition:CATalog?
Returns a comma-separated list of the currently defined converter types.
The length of the list can be queried using [SENSe:]CONVerter:DEFinition:
COUNt?.
Return values:
<Type>
Example:
SENSe:CONVerter:DEFinition:CATalog?
returns something like
'ZVA-Z110','ZVA-Z110E','ZVA-Z140', ...
Usage: Query only
Options: R&S ZNA-K8
Manual operation: See "Defined Converter Types (Tabs)" on page 905
[SENSe:]CONVerter:DEFinition:COUNt?
Returns how many converter types are currently defined (pre-defined or user-defined)
Return values:
<DefCount>
Usage: Query only
Options: R&S ZNA-K8
Manual operation: See "Defined Converter Types (Tabs)" on page 905
[SENSe:]CONVerter:DEFinition:DEFine <Type>
Creates a converter type with default properties.
SCPI command reference