1331User Manual 1178.6462.02 ─ 29
Setting parameters:
<Type> The name of the new converter type. Must be unique among the
existing ones (see [SENSe:]CONVerter:DEFinition:
CATalog?).
Example:
SENSe:CONVerter:DEFinition:DEFine 'My converter
type'
creates a new converter type named "My converter type".
SENSe:CONVerter:DEFinition:CATalog?
returns something like 'ZVA-Z110',...,'My converter
type'
Usage: Setting only
Options: R&S ZNA-K8
Manual operation: See "Type" on page 905
[SENSe:]CONVerter:DEFinition:DELete <Type>
Deletes the related converter type.
Only user-defined converter types can be deleted.
Setting parameters:
<Type> Name of the converter type to be deleted.
Usage: Setting only
Options: R&S ZNA-K8
Manual operation: See "Delete" on page 906
[SENSe:]CONVerter:DEFinition:FREQuency <Type>, <StartFreq>, <StopFreq>
[SENSe:]CONVerter:DEFinition:FREQuency? <Type>
Sets/queries the start and stop frequency of (the converter port of) the related con-
verter type.
Parameters:
<StartFreq> Default unit: Hz
<StopFreq> Default unit: Hz
Parameters for setting and query:
<Type> Name of the converter type, whose start and stop frequencies
are set/queried.
Options: R&S ZNA-K8
Manual operation: See "Converter Port: Start Freq./Stop Freq." on page 905