Command reference
R&S
®
ZNA
1332User Manual 1178.6462.02 ─ 29
[SENSe:]CONVerter:DEFinition:IFRequency <Type>, <IF>
[SENSe:]CONVerter:DEFinition:IFRequency? <Type>
Sets/queries the fixed intermediate frequency of the related converter type.
Parameters:
<IF> Intermediate frequency
Default unit: Hz
Parameters for setting and query:
<Type> Name of the converter type, whose IF is set/queried.
Options: R&S ZNA-K8
Manual operation: See "IF" on page 905
[SENSe:]CONVerter:DEFinition:LMUL
<Type>, <LocalMultiplicator>
[SENSe:]CONVerter:DEFinition:LMUL? <Type>
Sets/queries the LO multiplicator of the related converter type.
Parameters:
<LocalMultiplicator>
Parameters for setting and query:
<Type> Name of the converter type, whose LO multiplier is set/queried.
Options: R&S ZNA-K8
Manual operation: See "Local Port: Multiplicator" on page 906
[SENSe:]CONVerter:DEFinition:LPOWer <Type>, <LocalNomPower>
[SENSe:]CONVerter:DEFinition:LPOWer? <Type>
Sets/queries the preferred input power at the LO In port of the related converter type.
Parameters:
<LocalNomPower> Default unit: dBm
Parameters for setting and query:
<Type> Name of the converter type, whose preferred LO input power is
set/queried.
Options: R&S ZNA-K8
Manual operation: See "Local Port: Nominal Power" on page 906
[SENSe:]CONVerter:DEFinition:LPOWer:OFFSet <Type>,
<LocalNomPowerCalOffset>
[SENSe:]CONVerter:DEFinition:LPOWer:OFFSet? <Type>
Sets/queries the cal power offset of the source port that is connected to the LO In port,
when converters of the related type are configured.
SCPI command reference