Command reference
R&S
®
ZNA
1333User Manual 1178.6462.02 ─ 29
See SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet and
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet
Parameters:
<LocalNomPowerCalOffset>Default unit: dBm
Parameters for setting and query:
<Type> Name of the converter type, whose cal power offset is set/quer-
ied.
Options: R&S ZNA-K8
[SENSe:]CONVerter:DEFinition:MPOWer
<Type>, <MaxInputPower>
[SENSe:]CONVerter:DEFinition:MPOWer? <Type>
Sets/queries the maximum tolerable input power at the RF In port of the related con-
verter type.
Parameters:
<MaxInputPower> Default unit: dBm
Parameters for setting and query:
<Type> Name of the converter type, whose maximum RF input power is
set/queried.
Options: R&S ZNA-K8
Manual operation: See "Source: Converter Max. Power" on page 906
[SENSe:]CONVerter:DEFinition:SMUL <Type>, <SourceMultiplicator>
[SENSe:]CONVerter:DEFinition:SMUL? <Type>
Sets/queries the source multiplicator of the related converter type.
Parameters:
<SourceMultiplicator>
Parameters for setting and query:
<Type> Name of the converter type, whose source multiplier is set/quer-
ied.
Options: R&S ZNA-K8
Manual operation: See "Source: Multiplicator" on page 905
[SENSe:]CONVerter:DEFinition:SPOWer <Type>, <SourceNomPower>
[SENSe:]CONVerter:DEFinition:SPOWer? <Type>
Sets/queries the preferred input power at the RF In port of the related converter type.
Parameters:
<SourceNomPower> Default unit: dBm
SCPI command reference