Command reference
R&S
®
ZNA
1334User Manual 1178.6462.02 ─ 29
Parameters for setting and query:
<Type> Name of the converter type, whose preferred RF input power is
set/queried.
Options: R&S ZNA-K8
Manual operation: See "Source: Nominal Power" on page 905
[SENSe:]CONVerter:DEFinition:SPOWer:OFFSet <Type>,
<SourceNomPowerCalOffset>
[SENSe:]CONVerter:DEFinition:SPOWer:OFFSet? <Type>
Sets/queries the cal power offset of the source port that is connected to the RF In port,
when converters of the related type are configured.
See SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet and
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet
Parameters:
<SourceNomPowerCalOffset>Default unit: dBm
Parameters for setting and query:
<Type> Name of the converter type, whose cal power offset is set/quer-
ied.
Options: R&S ZNA-K8
[SENSe:]CONVerter<Port>:IFPort <IfInput>
Selects input connectors to be used for the IF meas and IF ref connection between the
converter and the analyzer port <Port>.
Suffix:
<Port>
.
Converter number = VNA port number
Parameters:
<IfInput> DRACcess | REAR
DRACcess
Use the (optional) direct access connectors at the front panel of
the R&S ZNA.
REAR
Use the (optional) direct IF access connectors IF Reference <i>
and IF Meas <i> at the rear panel of the R&S ZNA.
Options: R&S ZNA-K8
Manual operation: See "IF Input" on page 903
[SENSe:]CONVerter<Port>:IFRequency <IF>
Sets/queries the intermediate frequency for converter <Port>.
SCPI command reference