Command reference
R&S
®
ZNA
1377User Manual 1178.6462.02 ─ 29
ON (1)
Common sweep setup for all standards (no preparation phase
for each standard)
Note that noise figure calibrations require individual setups, and
hence this setting is ignored.
*RST: OFF
Manual operation: See "General>Same Sweep Setup for all Standards"
on page 591
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes <Boolean>
Toggles the Multiple Calibration TYpes mode that allows you to calibrate a subset of
the available channels using channel-specific ports and calibration types (see Chap-
ter 4.5.8, "Parallel calibration of multiple channels", on page 227).
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine. Disable the MCTYpes
mode after the calibrations were saved using [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected[:DUMMy]. The latter must be called only once, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>
SCPI command reference