Command reference
R&S
®
ZNA
1448User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTMLength> Default unit: m
Options: R&S ZNA-K20
Manual operation: See "Max. DUT Measure Delay / Max. El. Length / Max. Mech.
Length" on page 804
[SENSe<Ch>:]HARMonic:PERMittivity:DATA <DUTPermittivity>
Sets/gets the (relative) permittivity of the DUT.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTPermittivity>
Options: R&S ZNA-K20
Manual operation: See "Permittivity / Velocity Factor" on page 804
[SENSe<Ch>:]HARMonic:RTIMe:DATA <DUTRiseTime>
Sets/gets the minimum rise time the user wishes to measure on the DUT.
Use [SENSe<Ch>:]HARMonic:RTIMe:THReshold to select the underlying rise time
definition.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseTime> Default unit: s
Options: R&S ZNA-K20
Manual operation: See "Rise Time" on page 805
[SENSe<Ch>:]HARMonic:RTIMe:THReshold <DUTRiseThreshold>
Defines how the rise time that is set using [SENSe<Ch>:]HARMonic:RTIMe:DATA
shall be interpreted. Allows to select between rise time definitions 20%-80% (default)
and 10%-90%.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.