Command reference
R&S
®
ZNA
1453User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]NFIGure:IFConst.................................................................................... 1455
[SENSe<Ch>:]NFIGure:IFConst:STATe.......................................................................... 1455
[SENSe<Ch>:]NFIGure:NTIMe......................................................................................1456
[SENSe<Ch>:]NFIGure:RFICorr.................................................................................... 1456
[SENSe<Ch>:]NFIGure:TEATtenuator............................................................................1457
[SENSe<Ch>:]NFIGure:VIRejection[:STATe]................................................................... 1457
[SENSe<Ch>:]NFIGure:CALibration:GTIMe <DetectorTime>
Defines the detector time of the AVG detector which is used for the gain measurement
during a NF calibration in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime> Detector time for the AVG detector
Default unit: s
Example:
SENSe1:NFIGure:CALibration:GTIMe 10ms
Options: R&S ZNA-K30
Manual operation: See "Gain Det Meas Time" on page 595
[SENSe<Ch>:]NFIGure:CALibration:NTIMe <DetectorTime>
Defines the detector time of the RMS detector which is used for the noise measure-
ment during a NF calibration in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime> Detector time for the RMS detector
Default unit: s
Example:
SENSe1:NFIGure:CALibration:NTIMe 50ms
Options: R&S ZNA-K30
Manual operation: See "Noise Det Meas Time" on page 595
[SENSe<Ch>:]NFIGure:DEFine <receiving test port>, <driving test port>[, <further
port pairs>...]
[SENSe<Ch>:]NFIGure:DEFine?
The setting initiates a noise figure measurement in channel <Ch> and defines the
receive/drive test port pairs to be measured. The query returns the defined receive/
drive test port pairs, or an empty string, if no pairs are defined.
Suffix:
<Ch>
.
Channel number
SCPI command reference