EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1453

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1453User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]NFIGure:IFConst.................................................................................... 1455
[SENSe<Ch>:]NFIGure:IFConst:STATe.......................................................................... 1455
[SENSe<Ch>:]NFIGure:NTIMe......................................................................................1456
[SENSe<Ch>:]NFIGure:RFICorr.................................................................................... 1456
[SENSe<Ch>:]NFIGure:TEATtenuator............................................................................1457
[SENSe<Ch>:]NFIGure:VIRejection[:STATe]................................................................... 1457
[SENSe<Ch>:]NFIGure:CALibration:GTIMe <DetectorTime>
Defines the detector time of the AVG detector which is used for the gain measurement
during a NF calibration in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime> Detector time for the AVG detector
Default unit: s
Example:
SENSe1:NFIGure:CALibration:GTIMe 10ms
Options: R&S ZNA-K30
Manual operation: See "Gain Det Meas Time" on page 595
[SENSe<Ch>:]NFIGure:CALibration:NTIMe <DetectorTime>
Defines the detector time of the RMS detector which is used for the noise measure-
ment during a NF calibration in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime> Detector time for the RMS detector
Default unit: s
Example:
SENSe1:NFIGure:CALibration:NTIMe 50ms
Options: R&S ZNA-K30
Manual operation: See "Noise Det Meas Time" on page 595
[SENSe<Ch>:]NFIGure:DEFine <receiving test port>, <driving test port>[, <further
port pairs>...]
[SENSe<Ch>:]NFIGure:DEFine?
The setting initiates a noise figure measurement in channel <Ch> and defines the
receive/drive test port pairs to be measured. The query returns the defined receive/
drive test port pairs, or an empty string, if no pairs are defined.