Command reference
R&S
®
ZNA
1455User Manual 1178.6462.02 ─ 29
[SENSe<Ch>:]NFIGure:HARMonic[:MAXimumorder] <MaximumOrder>
Sets the sideband order of the sideband correction for a noise figure measurement in
channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<MaximumOrder> Sets the number of higher order sidebands which are measured
and corrected during calibration and measurement.
*RST: 3
Options: R&S ZNA-K30
Manual operation: See "Sideband Correction" on page 375
[SENSe<Ch>:]NFIGure:HARMonic[:STATe] <State>
Activates or deactivates the sideband correction for a noise figure measurement in
channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<State> Enables (1) or disables (0) the sideband correction of the noise
figure calibration and measurement.
*RST: 1
Options: R&S ZNA-K30
Manual operation: See "Sideband Correction" on page 375
[SENSe<Ch>:]NFIGure:IFConst <IFFreq>
If manual selection is activated ([SENSe<Ch>:]NFIGure:IFConst:STATe ON, this
command allows you to select an IF for niose figure measurements in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<IFFreq> Default unit: Hz
Options: R&S ZNA-K30
Manual operation: See "General > IF Frequency" on page 374
[SENSe<Ch>:]NFIGure:IFConst:STATe <Boolean>
Allows you to switch between automatic and manual IF selection for niose figure mea-
surements in channel <Ch>