Example:
SEGM1:INS:SEL DWEL
Select the meas. delay to determine the sweep time in a new
sweep segment no. 1.
SEGM1:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz and a meas. delay of 10 ms.
SEGM1:SWE:TIME?
Query the sweep time in the new segment.
Manual operation: See "Optional Columns" on page 557
[SENSe<Ch>:]SEGMent<Seg>:NFIGure:VIRejection[:STATe]
<Boolean>
For segmented sweeps in noise figure channel <Ch>, if [SENSe<Ch>:]SEGMent:
NFIGure:VIRejection:STATe:CONTrol is set to ON, this command allows you to
activate or deactivate virtual image rejection in sweep segment <Seg>.
Suffix:
<Ch>
.
Channel number
<Seg> Segment number
Parameters:
<Boolean> ON (1)
Virtual image rejection activated
OFF (0)
Virtual image rejection deactivated
Options: R&S ZNA-K30
Manual operation: See "Optional Columns" on page 557
[SENSe<Ch>:]SEGMent:NFIGure:VIRejection:STATe:CONTrol <Boolean>
For segmented sweeps in noise figure channels, this command defines whether virtual
image rejection can be activated or deactivated per segment or per channel.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Virtual image rejection can be de-/activated per segment (using
[SENSe<Ch>:]SEGMent<Seg>:NFIGure:VIRejection[:
STATe])
OFF (0)
Virtual image rejection can be de-/acitvated per channel (using
[SENSe<Ch>:]NFIGure:VIRejection[:STATe])
*RST: OFF
Options: R&S ZNA-K30