GUI reference
R&S
®
ZNA
346User Manual 1178.6462.02 ─ 29
Remote command: 
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:MEASure:SENDed
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:SDEFine:SENDed
S<out><in>
Selects one of the four elements of the standard 2-port S-parameters as a measured
quantity for the active trace.
Remote command: 
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
All S-Params
With the active logical ports distributed to m DUTs with n
1
, ..., n
m
 ports, respectively, the
analyzer creates n
1
2
 + ... + n
m
2
 diagrams and displays the full set of S-parameters, one
in each diagram. The diagrams are arranged as an n x n grid.
If "All S-Params" would result in more than 100 diagrams being created, it is disabled.
Remote command: 
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:DEFine:SGRoup
More S-Params...
Opens a dialog that allows to select an arbitrary S-parameter as measured quantity,
along with a suitable Detector. See Chapter 5.2.2.3, "More S-Params dialog",
on page 350.
Y/Z/k/μ-Parameters
Allows you to select other measurement results that are typically used to characterize
a linear DUT. All these results can be calculated from the DUT's S-parameters.
If entered manually, the logical port numbers must be expressed using the same num-
ber of digits. I.e. for <out>=10 and <in>=9 you can specify S<out>< in> as S1009 or
S010009, but not as S109.
Z <- S-Parameters ← Y/Z/k/μ-Parameters
Allows you to select a converted impedance as measured quantity.
Meas softtool