GUI reference
R&S
®
ZNA
378User Manual 1178.6462.02 ─ 29
More Information ← Result Settings
Opens a dialog that displays the calculated measurement time for the setup variants
that are possible with the relevant VNA hardware (Direct source monitor access,
Reverse Coupler Configuration, Internal low noise preamplifier), if equipped. The
resulting measurement times are based on the VNA characteristics, and the Meas Set-
tings and DUT Settings configured in the "Quickset" dialog. Modify these settings to
achieve a reasonable measurement time.
Driving Port
Allows you to evaluate the influence of the different front side receiver inputs and the
"Preamp. Gain" on page 527 "b1" ("Isolation Amplifier") on the measurement time.
Setting the "Receiver Input" and "Isolation Amplifier" here is equal to activating it in the
"Input/Output tab" on page 677 of the "Port Configuration" dialog or the Power tab of
the "Pwr Bw Avg" softtool, respectively.
In addition, it is possible to specify an external attenuation at the driving port to let the
"Quickset" logic choose the suitable source powers for calibration and measurement.
Remote command:
[SENSe<Ch>:]NFIGure:QSET:ATTSource