For SMARTerCals, an additional receiver power calibration at one of the ports is
required:
"Don't Show this Dialog Again" has the same effect as disabling "Show Cal Kit Label".
A green checkmark indicates that the calibration data of a standard has been acquired
successfully. A green checkmark after the port symbol indicates that the minimum
number of calibration measurements for the port has been performed.
Tip:
●
For a TRL calibration, at least one line standard must be measured between any
pair of calibrated ports. See Chapter 4.5.1.9, "TRL calibration", on page 195.
●
If the selected calibration kit comprises a Sliding Match, then for every required
Match measurement either the Match or at least three positions of the Sliding
Match must be measured. See Chapter 4.5.2.3, "Sliding Match standards",
on page 203.
Remote command:
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
See also: [SENSe<Ch>:]CORRection:COLLect:LOAD:SELected