Command reference
R&S
®
ZNA
973User Manual 1178.6462.02 ─ 29
CALCulate:CALValidate:RESPonse:PHASe:UPPer........................................................... 974
CALCulate:CALValidate:RESPonse:PRESet..................................................................... 974
CALCulate:CALValidate:RESPonse:REAL:LOWer.............................................................974
CALCulate:CALValidate:RESPonse:REAL:UPPer..............................................................974
CALCulate:CALValidate:RUN.......................................................................................... 974
CALCulate:CALValidate:RUN:RESult?............................................................................. 974
CALCulate:CALValidate:STANdard.................................................................................. 975
CALCulate:CALValidate:AVERage[:STATe] <Boolean>
Parameters:
<Boolean>
Manual operation: See "Preferences" on page 612
CALCulate:CALValidate:CHARacteriza <String>
Selects the CalU characterization to be used for the cal validation.
Use CALCulate:CALValidate:STANdard to select the CalU standard to be used
for the validation.
Parameters:
<String> *RST: 'Factory'
Manual operation: See "Characterization" on page 611
CALCulate:CALValidate:FORMat <Format>
Parameters:
<Format> MLINear | REAL
Manual operation: See "Preferences" on page 612
CALCulate:CALValidate:RESPonse:IMAGinary:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:IMAGinary:UPPer <Port>, <Value>
Parameters:
<Port>
<Value>
Manual operation: See "Validation Limits" on page 611
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer <Port>, <Value>
Defines the maximum magnitude deviations from the reference characterization S-
parameters that are allowed during cal validation.
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.