Command reference
R&S
®
ZNA
974User Manual 1178.6462.02 ─ 29
Parameters:
<Port> Physical analyzer port no.
<Value> *RST: -0.2 (LOWer) | +0.2 (UPPer)
Default unit: dB
Manual operation: See "Validation Limits" on page 611
CALCulate:CALValidate:RESPonse:PHASe:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:PHASe:UPPer <Port>, <Value>
Defines the maximum phase deviations from the reference characterization S-parame-
ters that are allowed during cal validation.
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.
Parameters:
<Port>
<Value> *RST: -2 (LOWer) | +2 (UPPer)
Default unit: degrees
Manual operation: See "Validation Limits" on page 611
CALCulate:CALValidate:RESPonse:PRESet
Usage:
Event
Manual operation: See "Validation Limits" on page 611
CALCulate:CALValidate:RESPonse:REAL:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:REAL:UPPer <Port>, <Value>
Parameters:
<Port>
<Value>
Manual operation: See "Validation Limits" on page 611
CALCulate:CALValidate:RUN
Runs the cal validation with its configured settings.
Usage:
Event
Manual operation: See "Validate" on page 610
CALCulate:CALValidate:RUN:RESult?
Displays the result (PASS/FAIL) of the preceding cal validation (CALCulate:
CALValidate:RUN).
SCPI command reference