Command reference
R&S
®
ZNA
1045User Manual 1178.6462.02 ─ 29
CALCulate:FMODel:ISD:PORT:SKIP:LEFT
CALCulate:FMODel:ISD:PORT:SKIP:NONE
CALCulate:FMODel:ISD:PORT:SKIP:RIGHt
Tells the ISD tool which ports (in the measured DUT + Test Fixture file) shall be skip-
ped when the tool is run.
●
...:LEFT: skip the ports on the left (according to the port order specified using
CALCulate:FMODel:ISD:PORT:ORDer)
●
...:NONE: do not skip any ports
●
...:RIGHt: skip the ports on the right (according to the port order specified using
CALCulate:FMODel:ISD<Ph_pt>:PORT:ORDer)
Equivalent to setting the ports_to_skip batch mode parameter of the ISD tool to -1,
0 or -2, respectively.
Use CALCulate:FMODel:ISD:PORT:SKIP to define an arbitrary set of ports to be
skipped.
Usage:
Event
Manual operation: See "Ports to Skip" on page 738
CALCulate:FMODel:ISD:PRESet
Restores the global default settings of the ISD tool.
Usage:
Event
Manual operation: See "Reset to Default" on page 730
CALCulate:FMODel:ISD:RUN:RUN
Runs the ISD tool.
Before executing this command, make sure that
●
the test coupon measurement (using CALCulate:FMODel:ISD:COUPon:
MEASure:OPEN and CALCulate:FMODel:ISD:COUPon:MEASure:SHORt for
1xOpen+1xShort, CALCulate:FMODel:ISD:COUPon:MEASure otherwise) fin-
ished successfully, or the test coupon data were successfully loaded from file
(using CALCulate:FMODel:ISD:COUPon:MEASure:FILename
●
the measurement of DUT + test fixture (using CALCulate:FMODel:ISD:DUT:
MEASure) finished successfully
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\ZNA\Embedding. If result files with the same name already exist, they will be
overwritten.
Example:
See Chapter 8.2.9, "Fixture modeling", on page 1762
Usage: Event
Manual operation: See "Run <Fixture Modeling Tool>" on page 734
SCPI command reference