Command reference
R&S
®
ZNA
1375User Manual 1178.6462.02 ─ 29
If several calibration units are connected, use SYSTem:COMMunicate:RDEVice:
AKAL:ADDRess to select a unit for the calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFOPort | POPTport | PFNPort |
VUOSm | CPUosm | UMDelay | PUMDelay
See [SENSe<Ch>:]CORRection:COLLect:AUTO:
CONFigure on page 1367.
<Characterization> Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
– The empty string (' ') refers to the factory calibration of (and
stored on) the active calibration unit.
– A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
– A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
– A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied.
<TestPort2> Second test port number of the analyzer
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Usage: Setting only
Manual operation: See "Apply" on page 582
[SENSe<Ch>:]CORRection:COLLect:AUTO:UTHRough <TestPort1>, <TestPort2>
With this command, you can perform the Unknown Through measurement during an
automatic UOSM calibration.
Although it works for all calibration units, it is only useful if the calibration unit selected
using SYSTem:COMMunicate:RDEVice:AKAL:ADDRess is an inline calibration sys-
tem (ICS).
Suffix:
<Ch>
.
Channel number
SCPI command reference