705User Manual 1178.6462.02 ─ 29
Phase Span ← Port-specific parameters (table)
Total phase difference between first and last phase sweep point (with constant phase
steps).
Remote command:
SOURce<Ch>:CMODe:PORT<Pt>:PHASe:SPAN
Gen / Channel Base Power / Port Power Offset / Power Result ← Port-specific
parameters (table)
Same functionality as in the Port Settings dialog. See:
"Source Gen" on page 671
System error corrections must be performed with alternating source signals ("Gen"
unchecked).
"Channel Base Power" on page 630
"Port Power Offset" on page 631
True Differential Mode
See "True Differential Mode" on page 353.
This button and the Imbalance Sweep Settings below it are only visible if option
R&S ZNA-K6 (True differential mode) is available, and is only enabled if Source Coher-
ence On is unchecked.
Remote command:
SOURce<Ch>:TDIF[:STATe]
Imbalance Sweep Settings ← True Differential Mode
These settings are only available if True Differential Mode is checked. They configure
the imbalance sweep types in true differenntial mode (see Chapter 4.7.5.2, "Amplitude
imbalance and phase imbalance sweep", on page 278). Use the Sweep Type tab to
activate one of them.
Compensate Imbalance ← Imbalance Sweep Settings ← True Differential Mode
Defines whether the imbalance of the DUT's balanced input line conductors shall be
compensated before starting an imbalance sweep (see "Imbalance Compensation of a-
Waves" on page 278
Remote command:
CALCulate<Ch>:TDIF:IMBalance:COMPensation[:STATe]
Phase > Port / Start Phase / Stop Phase ← Imbalance Sweep Settings ← True Dif-
ferential Mode
Configures the phase imbalance sweep type in true differential mode:
"Port" selects the balanced logical port to be swept
"Start Phase" and "Stop Phase" define the phase difference at the start/stop of
each sweep
Remote command:
SOURce<Ch>:TDIF:IMBalance:PHASe:LPORt
SOURce<Ch>:TDIF:IMBalance:PHASe:STARt
SOURce<Ch>:TDIF:IMBalance:PHASe:STOP