EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 822

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
822User Manual 1178.6462.02 ─ 29
An acoustic signal and a TTL signal indicating pass or fail can be generated in addi-
tion.
Remote command:
CALCulate<Chn>:EYE:MASK:STATe
CALCulate<Chn>:EYE:MASK:DATA?
CALCulate<Chn>:EYE:MASK:FAIL?
Mask Fail Beep
This checkbox determines whether the R&S ZNA should make an audible beep on
mask failures.
Note: The R&S ZNA does not have a built-in audio device and loudspeaker. To hear
sounds, connect a USB audio device to the instrument or operate it via remote desk-
top.
Remote command:
CALCulate<Chn>:EYE:MASK:FAIL:BEEP
Export Test Results
Opens a dialog that allows the user to save the mask test results to an ASCII file.
Remote command:
MMEMory:STORe:EYE:MASK:RESults
Applic softtool

Table of Contents