EasyManua.ls Logo

Philips Azurion - 6.14.2 Using SmartMask; 6.15 ECG Triggering

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 T
o adjust the transparency of the image, tap + or - on the touch screen module for the following
masks:
Vessel
Device
NOTE You can only adjust transparency when uoroscopy is not acve.
9 Start uoroscopy for the clinical procedure.
10 Insert the device when the subtracted vessel map is visible.
6.14.2 Using SmartMask
SmartMask allows you to use a previously acquired image as a vessel mask.
You can choose the image that you want to use for SmartMask. SmartMask images must have the same
projecon and source-to-image distance sengs as the current acquision sengs, but they can be
from a dierent series for the same paent.
1 Idenfy the series containing the desired image and open the series for review.
For more informaon, see the following secons:
Reviewing a Series using the Review Window (page 117)
Reviewing a Series using the Touch Screen Module (page 118)
2 Select the image for the vessel mask in the acquision window using the touch screen module or
the viewpad.
3 When the desired image is displayed in the live window, enable SmartMask by doing one of the
following:
On the control module, press SmartMask.
In the X-ray Sengs task on the touch screen module, select Roadmap and then select
SmartMask.
4 Start uoroscopy.
For more informaon, see Performing Fluoroscopy (page 81).
Aer subtracon, the current image is set as the SmartMask image.
5 Insert the device.
6.15 ECG Triggering
ECG triggering allows you to acquire images in the same phase of the heart cycle. The ECG signal is used
to generate the ECG trigger pulses with an adjustable delay.
To be able to start ECG triggered exposure or uoroscopy, establish a proper ECG signal. The system is
ready to start, but waits a limited period of me for an ECG signal. The system does not generate X-ray
before the ECG signal is recognized. ECG triggering is only applicable for uoroscopy and a limited set of
exposure procedures.
NOTE For Single Shot triggering, only one image is generated this way.
By default, ECG triggering is not acv
ated. Once acvated, the sengs remain valid unl switched o
or a new paent is selected. Selecng a procedure that is not supported by ECG triggering, for example,
Rotaon scan or Bolus Chase, automacally deacvates ECG triggering and the controls in the ECG
triggering task panel are unavailable.
Performing Procedures ECG Triggering
Azurion Release 1.2 Ins
trucons for Use 102 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals