Storing Fluoroscopy Series and Images
Y
ou can store acquired uoroscopy series and images in the paent’s le.
You can retrieve them in the Series task.
You can store individual images while you are performing uoroscopy, and store a uoroscopy series
aer acquiring it.
1 Start uoroscopy.
For more informaon, see Acquiring Fluoroscopy Images (page 80).
2 To store (or grab) individual images while you are performing uoroscopy, do one of the following:
On the control module, press and hold Fluoro Store.
On the touch screen module or in the acquision window, select and hold Fluoro Store.
Each image acquired while you hold the Fluor
o Store buon is stored. When you review the
images, the following symbol is displayed in the top right-hand corner of the image, indicang that
it is a stored image:
3 T
o store the series, do the following:
a Stop uoroscopy.
The last image in the acquired series is displayed as a Last Image Hold image.
b Do one of the following:
On the control module, press Fluoro Store.
On the touch screen module top bar or in the acquision window, select Fluoro Store.
The uor
oscopy series is stored. When you review the series, the following symbol is displayed in
the top right-hand corner of each image, indicang that it is a stored series:
Reseng the Fluor
oscopy Buzzer
When the cumulave uoroscopy me reaches 5 minutes, you are given an audible signal.
The indicator lights ash at the Reset Fluoroscopy Buzzer buons on the review module and the
control module, and a nocaon is displayed on the touch screen module.
NOTE Fluoroscopy is switched o automacally aer 10 minutes of uninterrupted uoroscopy.
1 To switch o the audible signal, do one of the following:
On the control module or on the review module, press Reset Fluoroscopy Buzzer.
On the touch screen module, tap Reset.
2 Connue with uor
oscopy if appropriate.
Performing Procedures Acquiring Images
Azurion Release 1.2 Ins
trucons for Use 82 Philips Healthcare 4522 203 52421