Philips Azurion - 10.8.4 Changing RVA Default Settings

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Item Default Sengs Input Range
LVA Monoplane Volume Methods List Area Length Not applicable
LVA Monoplane Regression formula EDV, ESV = 0.783, Vcalc = –3.759 Not applicable
LVA Biplane Volume Methods List Area Length Not applicable
LVA Biplane Regression formula EDV, ESV = 0.989, Vcalc = –8.1 Not applicable
LVA R
otaon/Angulaon Range –10 degrees to +10 degrees –20 degrees to +20 degrees
6 Click Sa
ve to save your changes.
10.8.4 Changing RVA Default Sengs
1 On the System menu, click Customizaon .
2 On the le side of the screen, in the Measurements and Analysis secon, click Right Ventricle
Analysis.
3 Change the following sengs, as desired:
Item Sengs
Default Index method BSA
BSA^1.219
Weight
Age Threshold Enter a value in the box to specify the child/adult age threshold.
Monoplane Volume Method List Pyramid
Monoplane Regression Formulas RAO30, EDV, ESV = 0.898, Vcalc = 3.862, Adults
Y
ou can select a predened regression formula, or you can enter a user-dened
formula in the boxes provided.
Biplane Volume Method List Area Length
Simpson
Biplane Regression Formulas Area Length
AP/Lateral, EDV, ESV = 0.779, Vcalc = –1.807, Adults
RAO30/LAO60, EDV, ESV = 0.79, Vcalc = 0.238, Adults
RAO45/LAO45, EDV, ESV = 0.737, Vcalc = –1.435, Adults
RAO60/LAO30, EDV, ESV = 0.749, Vcalc = 0.836, Adults
Any projecon, EDV, ESV = 0.76, Vcalc = –0.2, Adults
AP/Lateral, EDV, ESV = 0.898, Vcalc = 2.8, Children
AP/Lateral, EDV, ESV = 0.68, Vcalc = 0, Children
Simpson
AP/Lateral, EDV, ESV = 0.649, Vcalc = 0, Children
You can select a predened regression formula, or you can enter a user-dened
formula in the boxes provided.
Rotaon Range Enter a range in the boxes in which warnings are suppressed.
Angulaon Range Enter a range in the boxes in which warnings are suppressed.
4 T
o undo any changes that you have made in the Right Ventricle Analysis panel, click Undo Changes.
5 Alternavely, to restore the system sengs to default values, click Reset Default.
2D Quant
ave Analysis (Opon) 2D-QA Sengs
Azurion Release 1.2 Ins
trucons for Use 185 Philips Healthcare 4522 203 52421