Item Specicaon
Electrical safety measures ECG connector: The analogue output of c
ompable Physio/ECG monitoring
equipment may be connected to the X1 interface on the AD7X(NT) paent table.
When specied in the accompanying documentaon of the Physio/ECG equip-
ment, the funconal earth plug may be connected to one of the POAG type func-
onal ground pins on the AD7X(NT) paent table.
Injector connector: Compable contrast injector equipment may be connected
to the X2 interface on the AD7X(NT) paent table. When specied in the accom-
panying documentaon of the Physio/ECG equipment, the funconal earth plug
may be connected to one of the POAG type funconal ground pins on the
AD7X(NT) paent table.
Secondary Circuit Outlet: Supply mains connecon for external cered medical
equipment with a supply voltage rang of 230 Vac not exceeding an input power
rang of 600 VA. Connecon of external equipment to this interface is only al-
lowed if the equipment does not have other galvanic connecons to supply
mains grounds or building steel (such as unintended ground loops, for example,
using shielded cables shall not be allowed).
Foot switch (2x): Connecon of foot switch.
17.6 Con
tacng the Manufacturer
You can contact the manufacturer by post or by email.
Manufacturer's Address
Postal address Philips Medical Systems Nederland B.V.
V
eenpluis 4-6
5684 PC Best
The Netherlands
Email address healthcare@philips.com
Regulatory In
formaon Contacng the Manufacturer
Azurion Release 1.2 Ins
trucons for Use 348 Philips Healthcare 4522 203 52421