Item Specicaon
Electrical safety measures ECG connector: The analogue output of c
ompable Physio/ECG monitoring
equipment may be connected to the X1 interface on the AD7X(NT) paent table.
When specied in the accompanying documentaon of the Physio/ECG equip-
ment, the funconal earth plug may be connected to one of the POAG type func-
onal ground pins on the AD7X(NT) paent table.
Injector connector: Compable contrast injector equipment may be connected
to the X2 interface on the AD7X(NT) paent table. When specied in the accom-
panying documentaon of the Physio/ECG equipment, the funconal earth plug
may be connected to one of the POAG type funconal ground pins on the
AD7X(NT) paent table.
Secondary Circuit Outlet: Supply mains connecon for external cered medical
equipment with a supply voltage rang of 230 Vac not exceeding an input power
rang of 600 VA. Connecon of external equipment to this interface is only al-
lowed if the equipment does not have other galvanic connecons to supply
mains grounds or building steel (such as unintended ground loops, for example,
using shielded cables shall not be allowed).
Foot switch (2x): Connecon of foot switch.
17.6 Con
tacng the Manufacturer
You can contact the manufacturer by post or by email.
Manufacturer's Address
Postal address Philips Medical Systems Nederland B.V.
V
eenpluis 4-6
5684 PC Best
The Netherlands
Email address healthcare@philips.com
Regulatory In
formaon Contacng the Manufacturer
Azurion Release 1.2 Ins
trucons for Use 348 Philips Healthcare 4522 203 52421