1 Tube voltage (kV)
2 Tube current (mA)
Compability
The Cert
eray iX High Voltage generator is compable with the following tubes made by Philips Medical
Systems:
MRC 200+ 0407-ROT-GS 1004
MRC 200+ 0508-ROT-GS 1003
Labels
For informaon about labels, see Equipment Labels (page 325).
16.4 An-Scaer Grid
Detector Type Line Rate [lines/cm] Grid Rao Focal Spot Distance [cm]
FD12 N74 r14 f105
FD15 N70 r13 f100
FD20 N44 r12 f105
16.5 Mains Power
Acc
ording to the IEC60601-1 denion, the system is classied as class I equipment for connuous
operaon.
System
Conguraon Sengs
Recording mode of oper
aon Connuous
Supply conguraons 3 phase Y, 4 wires (L1, L2, L3, PE)
Mains voltage (± 10%), current, peak current, frequency 3~400 V, 76 A, 330 Apk, 50/60 Hz
3~415 V, 73 A, 325 Apk, 50/60 Hz
3~440 V, 69 A, 315 Apk, 50/60 Hz
3~480 V, 63 A, 300 Apk, 50/60 Hz
Maximum mains resistance 400 V: 140 mOhm (+ 80 mOhm internal resistance)
415 V: 215 mOhm (+ 80 mOhm internal resistance)
440 V: 325 mOhm (+ 80 mOhm internal resistance)
480 V: 465 mOhm (+ 80 mOhm internal resistance)
Maximum resistance at generator 400 V: 220 mOhm
415 V: 295 mOhm
440 V: 405 mOhm
480 V: 545 mOhm
Hospital mains fuse 125 amp with gG characteriscs (slow blow)
Maximum required power 110 kW
NOTE All connected phase wires shall have an upstream disconnect switch with a rang of up t
o
125 A.
NOTE All installaons and wiring up to the incoming mains power shall be installed and veried to
comply with applicable local regulaons.
Technical Informaon An-Scaer Grid
Azurion Release 1.2 Ins
trucons for Use 274 Philips Healthcare 4522 203 52421