c T
o close the dialog box without deleng the preset group, tap Cancel.
d To delete the preset group, tap Delete.
The preset group is deleted, including all presets contained within it.
8 To restore the factory default preset groups, do the following:
NOTE Restoring the factory default preset groups will overwrite all exisng presets and preset
groups, including customized presets and preset groups.
a Tap More and select Restore Defaults.
A conrmaon dialogue box is displayed.
b To close the dialog box without restoring the factory default preset groups, tap Cancel.
c To restore the factory default preset groups, tap Delete.
The factory default preset groups and presets are restored. Customized presets and preset
groups are deleted.
12.10 Changing Automac Posion Control Sengs
You can customize the Automac Posion Control (APC) sengs in the system for future use.
The system allows you to change, rename, copy, delete, and add new APC posions. You can also store
the current geometry for future use as an APC posion.
1 On the System menu, click Customizaon to display the System Customizaon window.
2 In the X-Ray Applicaon sengs group, click APC Posions.
If you are using a biplane system, the APC Posions dialog box allows you to select between
Monoplane or Biplane posions. You can use Monoplane posions on a biplane system, in which
case, only the posion informaon for the frontal channel is stored.
3 To rename an exisng APC posion, do the following:
a Select the APC posion in the Posion Name list.
The APC Posion Details are displayed.
b Enter a new name in the Posion Name eld.
The Posion Name list is automacally updated.
4 To change the sengs of a monoplane posion, do the following:
a Select Monoplane.
b Select the APC posion in the Posion Name list.
c Set the Rotaon Angle for the frontal stand using the slider or by entering a number in the box.
NOTE The labels used to indicate Rotaon Angle and Angulaon Angle depend on the
seng that is congured for Rotaon/Angulaon Display Flavor.
d Set the Angulaon Angle using the slider or by entering a number in the box.
e Set the Source Image Distance using the slider or by entering a number in the box.
f Select the Detector Orientaon from the drop-down list.
5 To change the sengs of a biplane posion, do the following:
User Cus
tomizaon Changing Automac Posion Control Sengs
Azurion Release 1.2 Ins
trucons for Use 223 Philips Healthcare 4522 203 52421