Vessel segment length accuracy is specied f
or distances up to 50 mm between user-dened markers
on an unforeshortened view of a vessel placed in the isocenter, using automac calibraon.
NOTE Using an inaccurate calibraon factor (due to, for example, foreshortening, inaccurate
posion of the calibraon object, or calibraon on a small diameter catheter) may lead to
addional errors in measured lengths and diameters.
QVA
QVA Analysis Results Accuracy (S
ystemac Error) Precision (Random Error)
Vessel diameter < 0.2 mm (for diameters ≤ 20 mm)
< 1% (for diameters > 20 mm)
< 0.2 mm
Vessel segment length < 1.0 mm < 2.0 mm
Vessel diameter accuracy is specied f
or measurements performed on a vessel placed in the isocenter,
using automac calibraon.
Vessel segment length accuracy is specied for distances up to 50 mm between user-dened markers
on an unforeshortened view of a vessel placed in the isocenter, using automac calibraon.
NOTE Using an inaccurate calibraon factor (due to, for example, foreshortening, inaccurate
posion of the calibraon object, or calibraon on a small diameter catheter) may lead to
addional errors in measured lengths and diameters.
References
Computaons in 2D-QA are performed according to methods described in medical literature.
Author Arcle
Reiber, J.H.C. et al. On-line quanc
aon of coronary angiograms with the DCI system. MedicaMundi, 34, no.
3, 1989. pp. 89-98.
Van der Zwet P.M.J. et al. A new approach for the automated denion of path lines in digized coronary angio-
grams. Int. J. Cardiac Imaging, 5, no. 2-3, 1990. pp. 75-83.
Van der Zwet, P.M.J. et al. An on-line system for the quantave analysis of coronary arterial segments. Computers in
Cardiology, 1990.
Austen, W.G. et al. A reporng system on paents evaluated for coronary artery disease. Report of the Ad Hoc
Commiee for grading of coronary artery disease. Council on Cardio-vascular Surgery,
American Heart Associaon. Circulaon 51, no. 2, 1975. pp. 7-40.
Reiber, J.H.C. et al. Assessment of dimensions and image quality of coronary contrast catheters from cine an-
giograms. Catheterizaon and Cardio-vascular Diagnosis, 11, 1985, pp. 521-531.
10.6 LVA / RVA
The L
VA and RVA applicaons have similar tasks, and they are described together in the following
secons.
Addional steps that are required for Biplane LVA and Biplane RVA are indicated where appropriate.
Le Ventricle Analysis (LVA)
You use LVA to establish end diastolic (ED) and end systolic (ES) contours of the le ventricle, to
determine ventricular volumes and wall moon. You can create, store and print reports of the analysis.
2D Quantave Analysis (Opon) LVA / RVA
Azurion Release 1.2 Ins
trucons for Use 167 Philips Healthcare 4522 203 52421