14.13 Showing the Monitor Test Image
To assist with maintenance, you can make the system display the Society of Moon Pictures and
Television Engineers (SMPTE) test image.
When the test image is displayed, you cannot use the system.
1 Ensure that the service applicaon is not in use and that a remote assistance session is not being
performed.
2 Click System in the review window and select Show Monitor Test Image.
A dialog box is displayed, requesng you to conrm that you want to display the test image.
3 Do one of the following:
• To close the dialog box without displaying the test image, click No.
• To display the test image, click Yes.
4 To stop displaying the image and restore the system to normal use, press any key or mouse buon.
14.14 XperCT Calibraon
NOTE The following procedure is fully automated. You do not need to enable or disable X-ray.
Interact with the system when instructed to do so by the system.
1 On the touch screen module, tap Tools.
2 In the Tools menu, tap Xper-CT Calibraon.
3 In the Xper-CT Calibraon submenu, tap one of the following opons, and then follow the guidance
on the touch screen module:
• Detector Prop
• Detector Roll
• Detector Prop + Roll
4 (FlexMove only) On the control module, use the Move Beam XY Motorized joysck to posion the
stand in the transverse zero posion.
You can also do this by selecng a procedure in which the stand moves automacally to the
transversal zero posion.
5 Press and hold the acquision hand switch or foot switch.
The system performs the calibraon procedure.
6 Release the acquision hand switch or foot switch when the system indicates that the calibraon
procedure is nished.
7 To stop the calibraon procedure while it is in progress, tap Abort.
8 Tap Close to return to the Tools menu.
14.14.1 Pre-Scan Calibraon
When you select an head procedure with XperCT, a pop-up message informs you to run the pre-scan
calibraon procedure for beer image quality. The pop-up message will not be displayed for the next
hour.
Maintenance Showing the Monitor Test Image
Azurion Release 1.2 Ins