Analysis Results Descripon Formula (if applicable)
Heart Rate (BPM) The Heart Rate is indicated in Beats Per Minute.
Body Surface Area (BSA) Body Surface Area is calculated from the paen
t’s height and weight. BSA can be
used to generate indexed results.
Index Method The index method used to calculate indexed results.
Volume Method The selected volume method.
ED Volume Regression The formula used with the volume method to calculate the ED volume.
ES Volume Regression The formula used with the volume method to calculate the ES volume.
Contour Correcon Indicates whether contours were manually corrected during analysis.
Calibraon Object The calibraon method used and the size of the calibraon object.
Calibraon Factor The calibraon factor calculated by the system using the inputs in the Calibraon
task.
Series The series number of the series used for analysis.
ED Image The image number of the image used as the selected ED image.
ES Image The image number of the image used as the selected ES image.
Projecon (Frontal/Lateral for biplane
systems)
The projecon used during acquison (RAO/LAO).
ED Volume (EDV) The ED volume is calculated from using the volume method and the ED regres-
sion method.
The Indexed ED Volume is displayed if the paent’s demographics are available.
ES Volume (ESV) The ES volume is calculated from using the volume method and the ES regression
method.
The Indexed ES Volume is displayed if the paent’s demographics are available.
Stroke Volume (SV) The Stroke Volume is calculated as the
dierence of the ED Volume and the
ES Volume.
The Indexed StrokeVolume is displayed
if the paent’s demographics are avail-
able.
SV (ml) = EDV - ESV
Indexed SV (ml/m
2
) = SV ÷ BSA
Seng the P
aent Demographics
Some analysis results depend on correctly dened paent demographics, such as the paent’s height,
weight, and heart rate.
The paent’s height and weight allows the Body Surface Area (BSA) to be calculated, which in turn
allows indexed analysis results to be calculated. When available, the paent’s height and weight are
automacally retrieved from the paent database, otherwise you can enter them manually.
The paent’s heart rate allows the Cardiac Output and the Cardiac Index to be calculated. The paent’s
heart rate is automacally entered if this informaon is available in the paent database, or you can
enter the informaon manually.
You can edit the acquision paent’s demographics using the following procedure.
1 In the control panel, click Edit Paent Demographics.
2 If the paent’s height and weight informaon is not displayed, or if it is incorrect, enter the correct
informaon.
3 Enter the paent’s heart rate.
4 Click OK to close the dialog box and return to the Result task.