1 On the t
ouch screen module, tap Tools.
2 In the Tools menu, tap Xper-CT Calibraon.
3 In the Xper-CT Calibraon submenu, tap Pre-scan and follow the guidance on the touch screen
module.
4 Press and hold the acquision hand switch or foot switch.
The system performs the calibraon procedure.
5 Release the acquision hand switch or foot switch when the system indicates that the calibraon
procedure is nished.
6 To stop the calibraon procedure while it is in progress, tap Abort.
7 Tap Close to return to the Tools menu.
14.15 Environmental Impact of the System
You can assess the environmental impact of the system by measuring the typical energy consumpon
during various operaonal modes.
To reduce the environmental impact of the system, switch it o when it is not in use. However, bear in
mind any clinical limitaons that might make switching the system o impraccal.
For informaon, visit the following website:
www.cocir.org/index.php?id=198
14.16 Disposing of the System
Philips Healthcare is concerned about protecng the natural environment and ensuring connued safe
and eecve use of the system through proper support, maintenance and training.
Philips Medical Systems equipment is designed and manufactured to comply with relevant guidelines
for environmental protecon. As long as the system is properly operated and maintained it presents no
risk to the environment. However, the equipment may contain materials which could be harmful to the
environment if disposed of incorrectly. Use of such materials is essenal for the implementaon of
certain funcons and for meeng certain statutory and other requirements.
Final disposal is when the responsible or
ganizaon disposes of the equipment or system in such a way
that it can no longer be used for its intended purpose.
NOTE Computer disks that are part of the system could contain personal data. These disks should
be disposed of according the service instrucons.
Do not dispose of the system, or any parts of it, with industrial or domesc waste. The system may
contain materials such as lead, tungsten or oil, or other hazardous substances that can cause serious
environmental polluon. The system also contains privacy sensive informaon which should be
properly erased.
Philips Medical Systems can support you with recovery of reusable parts, recycling useful materials, and
the safe and eecve disposal of equipment.
For more informaon about recycling Philips Medical Systems products, refer to the following website:
www.medical.philips.com/main/about/sustainability/recycling/index.wpd
Maintenance Environmental Impact of the System
Azurion Release 1.2 Ins
trucons for Use 264 Philips Healthcare 4522 203 52421