Task Frequency Required Personnel
Check the mechanical x
aon of the monitor ceiling sus-
pension
Every year Technician
Clean the top side of the FlexMove carriage Every year Technician
Check the ECG and injector relays Every year Technician
Perform mechanical maintenance of the paent table Every 4 years Technician
Check the electrical safety Every 2 years Technician
Check the X-ray safety Every 2 years Technician
Check the X-ray protecon devices Every year Technician
14.5 User Quality Control Mode
T
o enable X-ray dose-related constancy tesng, the equipment provides a User Quality Control Mode
(UQCM) to perform X-ray dose-related tests.
UQCM is intended for trained hospital radiaon physicists or service engineers and comprises special
user quality control procedures that are accessible by using a service dongle. This dongle is only made
available by Philips Medical Systems once the user has followed the appropriate training. For more
informaon, see Contacng the Manufacturer (page 348).
In case of failure of the measurements performed under UQCM, contact technical support.
14.6 User Vericaon Test
Perform this procedure to verify system funconality.
1 Test the collimator funcons using uoroscopy and conrm that the X-ray on indicators are on while
X-ray is acve.
For more informaon about X-ray on indicators, see X‐ray On Indicators (page 77).
2 Test the table and the stand movements without using X-ray.
3 Perform the following test using a user-dened phantom for constancy evaluaon.
a Posion the stand in a vercal posion.
b Place the phantom on the table and in the X-ray beam.
c For constancy evaluaon, use a xed source-to-image distance and a consistent choice for eld
of view.
d Perform uoroscopy and check whether the X-ray indicaons are as expected, and that the kV
and mA values are within the expected ranges for the constancy evaluaon.
e Perform a digital cardiac or vascular exposure run and check whether the X-ray indicaons are
as expected, and the kV and mA values are within the expected ranges for the constancy
evaluaon.
Maintenance User Quality Control Mode
Azurion Release 1.2 Ins
trucons for Use 256 Philips Healthcare 4522 203 52421