EasyManua.ls Logo

Philips Azurion - 10.3 Starting 2 D Quantitative Analysis; 10.4 Calibration Guidelines

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Guidance for QCA and QVA
Prevent foreshortening of objects by using projecons where the object under analysis is in a plane
parallel to the image detector.
Avoid imaging with strong noise, background structures, or overlapping vessels.
Avoid imaging at 50/60 fps as the decreased resoluon in these images aects the accuracy of the
results.
Guidance for LVA and RVA
Use an acquision speed of at least 15 fps to allow selecon of images from non-ectopic beats and
in a proper end-diastolic and end-systolic phase.
Acquire images from angles as prescribed for the various volume and wall moon methods.
Instruct the paent in the use of breath-holding techniques for acquision of images for wall
moon analysis.
10.3 Starng 2D Quantave Analysis
1 Click the Processing task to display the image processing tools.
2 Click Measurements to display the Measurements task panel.
3 In the Open Analysis Tool secon, click the desired analysis tool buon to start the analysis.
Quantave Coronary Analysis
Quantave Vascular Analysis
Le Ventricular Analysis
Biplane Le Ventricular Analysis
Right Ventricular Analysis
Biplane Right Ventricular Analysis
NOTE To open only the frontal image or the lateral image of a biplane series in an analysis
applic
aon, right-click the image, point to Open With, and then click a monoplane
applicaon.
10.4 Calibraon Guidelines
A projecon of an anatomical object on an X-ray detector is geometrically magnied. If you want to
perform a realisc measurement in the corresponding X-ray image, you have to compensate for that
magnicaon. This is done through performing a calibraon on the X-ray image, and determining a
calibraon factor (CF) in units of millimeter/pixel.
There are two main types of calibraon:
2D Quant
ave Analysis (Opon) Starng 2D Quantave Analysis
Azurion Release 1.2 Ins
trucons for Use 158 Philips Healthcare 4522 203 52421

Table of Contents

Related product manuals