C12/F12 System
Dose r
ao (%) 100 110 130 155 185
Field Size (cm) 30 27 22 19 15
X-ray Protocol Paent Type Ref. AK (mGy/img)
Cerebral Cerebral 2 fps Low Default 1.021 1.132 1.358 1.650 2.015
The measurement c
ondions used are in accordance with Reference Air Kerma Measurement
Setup (page 311).
The following example gives the reference air kerma values for a vascular procedure for dierent eld
size values on F15 systems.
F15 System
Dose r
ao (%) 100 110 130 150 180 215 260
Field Size (cm) 39 37 31 27 22 19 15
X-ray Protocol Paent Type Ref. AK (mGy/img)
Head Clarity 2 fps Low Default 0.590 0.649 0.780 0.914 1.112 1.357 1.672
The measurement c
ondions used are in accordance with Reference Air Kerma Measurement
Setup (page 311).
The following example gives the reference air kerma values for a vascular procedure for dierent eld
size values on C20/F20 systems.
C20/F20 System
Dose r
ao (%) 100 130 145 170 200 240 280 330
Field Size (cm) 48 42 37 31 27 22 19 15
X-ray Protocol Paent Type Ref. AK (mGy/img)
Cere-
bral
Cere-
bral
2 fps
Low
Default 0.463 0.605 0.677 0.798 0.946 1.149 1.358 1.620
The measurement c
ondions used are in accordance with Reference Air Kerma Measurement
Setup (page 311).
NOTE The dose rao numbers may dier per procedure and per uoroscopy avor.
Mulphase Sengs
The vascular exposure procedure contains the default duraon and frame speed per phase.
For these procedures it is possible to manually change the frame speed and duraon per phase. For
more informaon about changing frame speed and duraon, see Changing Mulphase Acquision
Sengs (page 97).
The reference air kerma is dened per image and will not change at dierent frame speeds. However,
the cumulave skin dose is directly related to the frame speed and so, if the frame speed in one phase
is reduced by 50%, the cumulave skin dose in that phase is also reduced by 50%.
Conclusion: Consider lowering the frame speed, if possible.
Shuers and Wedges
When you apply proper collimaon, direct irradiaon of body parts not needed for the procedure is
prevented.
This reduces the dose area product and the sta dose, although the reference air kerma and the (peak)
skin dose are not inuenced.
In general, for example, 25% collimaon of the irradiated area will reduce the dose area product by
25%.
Technical Informaon System Sengs Inuencing the Radiaon Dose
Azurion Release 1.2 Ins
trucons for Use 292 Philips Healthcare 4522 203 52421