3 Open the paen
t database.
4 Click the device that you want to import from.
If the device is password-protected, enter the password in the dialog box displayed and click
Unlock.
A paen
t list is displayed showing the available studies from the selected device.
5 Select the desired paent study in the list.
The study details are displayed below the paent list.
6 To import the whole study, click Import.
A dialog box is displayed requesng you to conrm your acon.
7 Conrm your import by doing the following:
Click Link to import the data and merge paent details. Use this opon if the data that you are
imporng belongs to a paent who already has studies on the system.
Click Import to import the data without merging paent details.
Click Cancel to cancel the import.
If you click Link, a further dialog box is displayed. Check that the paent details are correct and then
click Link Data to import the data and merge the paent details. Alternavely, click Cancel to close
the dialog box without imporng the data.
8 To import a series from the study, click the Series tab.
The available series are displayed. No previews are shown in the pictorials as the series are not
stored in the local paent database.
9 Select the series that you want to import and click Import.
NOTE Do not remove the USB device, the CD, or the DVD unl the import process is complete
(the progress of the import process is displayed).
When the import process is complete, a preview image is displayed in the pictorial.
10 If you want to cancel the import process, click Stop.
7.8 Bolus Chase Reconstrucon
Bolus Chase Reconstrucon is a soware tool that is available as an opon on the system. It creates an
overview image of the arteries in the paents legs by automacally stching together successive
images acquired with the Flexible Dynamic Peripheral Angiography (FDPA) acquision protocol or Bolus
Chase acquision protocol.
The overview image is intended to assist you in viewing the original images. The overview image is not
intended to be used for diagnosis. Actual diagnosis (for example, analysis of occlusions) should be
based on informaon contained in the original X-ray images.
For informaon about acquiring X-ray images, see Bolus Chase (page 98).
NOTE Imported images cannot be used to create an overview image in Bolus Chase Reconstrucon.
The Bolus Chase Reconstrucon screen provides the following views and task panels.
Reviewing Bolus Chase R
econstrucon
Azurion Release 1.2 Ins
trucons for Use 123 Philips Healthcare 4522 203 52421