System X-ray Protocol Ref. AK (mGy/img)
C12/F12 with ClarityIQ (op-
on)
Cardio Pediatric <40 kg Cine 15 fps Low 0.019
Cardio Pediatric >40 kg Cine 15 fps Low 0.037
Cardiac Le Cor
onary 15 fps Low 0.037
Cerebral Cerebral 2 fps Low 1.021
Cerebral Cerebral 4 fps Low 1.020
Thorax Lungs 3 fps 1.259
Measurement c
ondions: paent type: default, eld size: 30 cm. All other sengs in accordance with
Reference Air Kerma Measurement Setup (page 311).
The following examples give the reference air kerma rate for the three uoroscopy avors for a typical
cardiac X-ray protocol.
System X-ray Protocol Flavor Ref. AK-rate (mGy/s)
C12/F12 Pediatric Low 0.183
Pediatric Normal 0.434
Pediatric High 0.684
Cardiac Low 0.238
Cardiac Normal 0.517
Cardiac High 0.719
C12/F12 with ClarityIQ (opon) Cardio Pediatric Low 0.061
Cardio Pediatric Medium 0.092
Cardio Pediatric Normal 0.144
Cardiac Low 0.149
Cardiac Medium 0.221
Cardiac Normal 0.518
Measurement c
ondions: paent type: default, eld size: 30 cm. All other sengs in accordance with
Reference Air Kerma Measurement Setup (page 311).
For an overview of a number of frequently used exposure procedures and uoroscopy avors, under
dened measurement condions, see Typical Reference Air Kerma (Rate) Values (page 295).
Paent Type
Although the system has an automac dose control mechanism that compensates for the various
depths of irradiated ssue, in some cases the image quality needs to be improved for very obese or
very thin paents. This is achieved by the system by removing or adding spectral ltering.
The paent type selecon may have an eect on the resulng reference air kerma. For opmal image
quality, you should select a paent type that matches the actual paent thickness. You can change the
paent type by eding the scheduled study. For more informaon about eding study details, see
Eding a Scheduled Study (page 51).
You can select one of the following paent types:
Paent T
ype Weight
Neonate <5 kg
Infant 5 - 15 kg
Child 15 - 40 kg
Small Adult 40 - 55 kg
Normal Adult 55 - 70 kg
Technical In
formaon System Sengs Inuencing the Radiaon Dose
Azurion Release 1.2 Ins
trucons for Use 290 Philips Healthcare 4522 203 52421