Right Ventricle Analysis (RVA)
Y
ou use RVA to establish end diastolic (ED) and end systolic (ES) contours of the right ventricle, to
determine ventricular volumes and wall moon. You can create, store and print reports of the analysis.
10.6.1 LVA / RVA Tasks
A set of predened tasks is used to ensure le or right ventricle analysis is performed in a logical way.
The LVA and RVA applicaons provide the following tasks in order:
Select Series
Calibraon
End Diastole
End Systole
Result
Aer you select a series, the Calibraon task is automacally opened.
Aer you complete calibraon, the End Diastole task is automacally opened.
NOTE Auto calibraon is available if the appropriate image aributes in the selected series (source
image distance, source object distance, and image plane pixel spacing) did not change during
acquision. If you choose auto calibraon in this case, ensure that the region of interest is in
the isocenter.
10.6.2 Select Series Task
You use the Select Series Task to select a series for analysis.
NOTE Only XA exposure images can be used for analysis.
NOTE Series with characteriscs outside the following ranges are subopmal for analysis:
Series with image pixel sizes greater than 1 mm.
Series with frame rate less than 15 fps.
Series captured with angulaon and rotaon angles that do match the angle
requirements for the selected volume method/regression formula.
1 Click Select Series in the tasks panel.
2 Select the desired image series in the Select Series dialog box and click Select to open the series.
10.6.3 Calibraon Task
To allow accurate measurement during analysis and to ensure that measurements are displayed in
relevant units, the image must be calibrated.
NOTE You can c
ongure default sengs for calibraon using the Customizaon screen. For details,
see Changing Default Calibraon Sengs (page 183).
You can perform c
alibraon automacally using Auto Calibraon, or manually using the Calibraon
task.
If you are only interested in calculang the ejecon fracon, you can skip calibraon for monoplane LVA
and monoplane RVA.
Condions
For manual calibraon, follow these guidelines:
2D Quant
ave Analysis (Opon) LVA / RVA
Azurion Release 1.2 Ins
trucons for Use 168 Philips Healthcare 4522 203 52421