Each X-ray protocol is associated with a de
ned list of automac posion control sengs. You can
change which posions can be recalled for each X-ray protocol.
NOTE Before you can select an automac posion control posion, it must exist in the list of
available posions.
For more informaon about managing automac posion control posions, see Changing Automac
Posion Control Sengs (page 223).
1 On the System menu, click Customizaon to display the System Customizaon window.
2 In the X-Ray Applicaon sengs group, click X-ray Protocols.
3 Select the desired X-ray protocol in the X-ray Protocols list.
The list displays the parent X-ray protocols by default. You can expand each parent protocol to allow
you to select a child protocol.
Changing the available posions for a parent protocol will make the selected posions available for
all the child protocols contained within the parent protocol.
Changing the available posions for a child protocol will make the selected posions available only
in the child protocol.
4 In the Details area, select the APC Posions you want to be available for the selected X-ray
protocol.
5 To change the order the selected posions are displayed in the system, do the following:
a In the APC Posions Order list, select the posion you want to move.
b Click the up or down buons to move the posion up or down in the list.
6 T
o undo any changes you have made, click Undo Changes.
7 To save your changes, click Save.
NOTE The changes take eect aer the next system shutdown and startup.
8 To close the System Customizaon window, click Close.
12.12 Changing Viewing Preferences
You can change some viewing sengs to suit the way you use the system.
The viewing sengs you can change are:
The X-ray image displayed when you open a series.
The way that navigaon and replay is managed between series.
Maximum series and study replay mes.
The way that angles are displayed on the system.
1 On the System menu, click Customizaon to display the System Customizaon window.
2 In the X-Ray Applicaon sengs group, click Viewing and Processing.
3 To change the default image displayed when you open a series, select a new seng in the Default
Acve X-ray Image list.
The sengs available are:
User Cus
tomizaon Changing Viewing Preferences
Azurion Release 1.2 Ins
trucons for Use 225 Philips Healthcare 4522 203 52421