1 P
osion the C-arm in the desired work posion.
2 Posion the region of interest in the isocenter.
For more informaon, see Isocentering (page 90).
3 Select the X-ray Sengs task on the touch screen module.
4 Tap Rotaonal Scan.
Step 1 Sengs of the rotaonal scan wizard is displayed.
5 From the control module, posion the detector in landscape or portrait posion.
6 Select the maximum source-to-image distance.
7 Select the desired eld of view.
8 Clear all objects from the rotaonal arc of the C-arm.
9 If you are performing a cardiac rotaonal scan, select the desired rotaon seng.
10 Switch Injector Coupling on or o as desired.
a If you switch Injector Coupling on, set the X-ray delay me.
11 Tap Next on the touch screen module.
Step 2 End posion of the rotaonal scan wizard is displayed.
12 To set the end posion, do one of the following:
Press Accept on the control module to move to the default end posion.
Posion the C-arm using the control module and select Next to set the end posion.
Step 3 Start posion of the rotaonal scan wizard is displayed.
13 To set the start posion, do one of the following:
Press Accept on the control module to move to the default start posion.
Posion the C-arm using the control module and select Next to set the start posion.
14 Tap Done to close the wizard.
15 If injector coupling is switched on, prepare the injector.
16 Start acquision using the exposure hand switch or the exposure foot switch pedal, and hold unl
the exposure stops.
NOTE BodyGuard is disabled during ac
quision.
6.16.3 XperCT
The XperCT procedure consists of a rotaonal scan. The acquired images are automacally sent to the
Intervenonal Tools workstaon.
The XperCT procedure is available only on systems ed with the 20-inch at detector.
For biplane systems, XperCT is only available on the frontal channel.
For informaon about XperCT calibraon, see XperCT Calibraon (page 263).
To be able to use XperCT, the table should be posioned within the following ranges:
Performing Procedures Rotaon Sc
ans
Azurion Release 1.2 Ins
trucons for Use 105 Philips Healthcare 4522 203 52421