1 P
osion the C-arm in the desired work posion.
2 Posion the region of interest in the isocenter.
For more informaon, see Isocentering (page 90).
3 Select the X-ray Sengs task on the touch screen module.
4 Tap Rotaonal Scan.
Step 1 Sengs of the rotaonal scan wizard is displayed.
5 From the control module, posion the detector in landscape or portrait posion.
6 Select the maximum source-to-image distance.
7 Select the desired eld of view.
8 Clear all objects from the rotaonal arc of the C-arm.
9 If you are performing a cardiac rotaonal scan, select the desired rotaon seng.
10 Switch Injector Coupling on or o as desired.
a If you switch Injector Coupling on, set the X-ray delay me.
11 Tap Next on the touch screen module.
Step 2 End posion of the rotaonal scan wizard is displayed.
12 To set the end posion, do one of the following:
• Press Accept on the control module to move to the default end posion.
• Posion the C-arm using the control module and select Next to set the end posion.
Step 3 Start posion of the rotaonal scan wizard is displayed.
13 To set the start posion, do one of the following:
• Press Accept on the control module to move to the default start posion.
• Posion the C-arm using the control module and select Next to set the start posion.
14 Tap Done to close the wizard.
15 If injector coupling is switched on, prepare the injector.
16 Start acquision using the exposure hand switch or the exposure foot switch pedal, and hold unl
the exposure stops.
NOTE BodyGuard is disabled during ac
quision.
6.16.3 XperCT
The XperCT procedure consists of a rotaonal scan. The acquired images are automacally sent to the
Intervenonal Tools workstaon.
The XperCT procedure is available only on systems ed with the 20-inch at detector.
For biplane systems, XperCT is only available on the frontal channel.
For informaon about XperCT calibraon, see XperCT Calibraon (page 263).
To be able to use XperCT, the table should be posioned within the following ranges:
Performing Procedures Rotaon Sc
ans
Azurion Release 1.2 Ins
trucons for Use 105 Philips Healthcare 4522 203 52421