The users details are displayed in the De
tails secon.
2 Click Reset Password.
A dialog box is displayed.
3 Enter a New Password.
You must follow these rules when seng a password:
The password eld cannot be empty.
Passwords cannot contain user names.
Passwords must conform to the password policy sengs (see Managing Users and System
Logon (page 230)).
If password complexity is enabled, passwords must contain characters from three of the
following categories:
Uppercase leers
Lower case leers
Numbers (0 through 9)
Non-alphabec characters (for example: ! $ # %)
4 Enter the same password in Conrm Password.
NOTE The password entered in Conrm Password must match the password entered in New
Password.
5 Do one of the following:
a To close the dialog box without reseng the user's password, click Cancel.
b To close the dialog box and reset the user's password, click Apply.
13.4 Changing General Paent and Workow Sengs
You can customize general workow sengs and specify sizes for the dierent paent types.
If the system local storage is full, the system automacally deletes data that is not protected to make
space available for newly acquired images. You can congure the system to protect every study upon
compleon.
You can congure the system to automacally start procedures that have been provided from XperIM.
You can simplify the DICOM workow to automacally mark all procedures as completed, and to
automacally produce a dose report when a procedure is closed.
You can change these basic paent and workow sengs:
Prevenng automac study deleon
Enabling a simplied DICOM workow
Enabling automac dose reporng
Neonatal, infant and child age limits
Adult circumference limits
Default paent type
Enabling support for Chinese, Japanese, and Korean (CJK) ideographic characters
Making the system compliant with requirements from the United States Department of Veterans
Aairs (VA)
1 On the System menu, click Customizaon to display the System Customizaon window.
2 In the General sengs group, click Workow.
3 To protect every study upon compleon, select the Prevent Automac Study Deleon check box.
System Adminis
traon Changing General Paent and Workow Sengs
Azurion Release 1.2 Ins
trucons for Use 232 Philips Healthcare 4522 203 52421