• DAP-meter: < 0.5 mm Al.
•
Wedge lter: 1 mm brass (CuZn37 R-019; 22 mm Al equivalent at 75 kV/HVL 3.5 mm Al)
Depending on the selected procedure, an addional lter may also be applied by the system, with the
following values (for beam liming devices with idencaon number 9896 010 22xxx):
Addional Filt
er Number Filter Filtraon
1 0.1 mm Cu + 1.0 mm Al 4.0 mm Al (75 kV/HVL 3.5 mm Al)
2 0.4 mm Cu + 1.0 mm Al 11.0 mm Al (75 kV/HVL 3.5 mm Al)
3 0.9 mm Cu + 1.0 mm Al 21.5 mm Al (75 kV/HVL 3.5 mm Al)
The following table shows air kerma values as a percentage of the curves as a funcon of the addional
lt
er selecon.
kV Filter
Air Kerma Value (%)
No pr
otecon
0.5 mm lead
equivalence
1.0 mm lead
equivalence
1.5 mm lead
equivalence
110 0 100 100 100 100
1 66 87 85 76
2 38 69 64 53
3 19 49 47 32
90 0 64 33 34 46
1 39 27 24 31
2 19 20 15 17
3 8.5 12 9.0 9.2
70 0 35 6.0 13 22
1 18 4.1 7.2 12
2 7.1 2.4 3.1 4.6
3 2.3 1.3 1.0 1.9
The following table shows normalized air kerma values as a percentage of the curves as a funcon of
the addional lt
er selecon.
Addional Filt
er
Normalized Air Kerma Val-
ue (%)
0 100
1 120
2 135
3 150
16.21.6 User Dose and Imaging In
formaon for Cone Beam CT Reconstrucons
This secon provides informaon about cone beam CT reconstrucons.
Phantoms and Measurement Methods
Dose Phantoms
The CT Dosimetry Phantom is the phantom used to determine the dose delivered during a Cone Beam
CT acquision. The phantoms are circular cylinders of polymethyl methacrylate and are 15 cm long.
Their density is 1.19 ± 0.01 g/cc. The phantom for tesng CT imaging of the body has a diameter of 32.0
cm, and the phantom for the head has a diameter of 16.0 cm.
Technical Informaon Protecon Against Stray Radiaon
Azurion Release 1.2 Ins
trucons for Use 318 Philips Healthcare 4522 203 52421