Errors in the c
alibraon factor propagate proporonally into QCA/QVA distance measurement. Relave
errors mulply with a factor of approximately 2 to 3 in the LVA/RVA computaons of absolute ventricle
volumes. Ejecon fracon, however, is not aected by these calibraon inaccuracies.
Guidance for Manual Catheter Calibraon
• Use a radiopaque catheter.
• Use a lled catheter to improve detecon and accuracy.
• Philips Medical Systems does not recommend catheter calibraon on empty catheters or catheters
below 6 French, as this may lead to an inaccurate calibraon factor. The error can be 20% or more.
2D-QA does not support catheters below 4 French.
• To improve accuracy, avoid low dose and high frame rates.
Guidance for Manual Sphere Calibraon
It is possible to use two dierent series for imaging the sphere and the anatomy under invesgaon.
However, ensure that the sphere and the anatomy have the same geometrical magnicaon in the X-
ray image. This means that the images are acquired with the following characteriscs:
• The same X-ray focus-object distance and same object-detector distance.
• The same angulaon and rotaon angles.
• The same table height.
Checking calibraon accuracy for your preferred catheter
1 Posion a catheter and a ruler close together, acquire images, and then perform catheter
calibraon.
2 In the X-ray image perform a QCA length measurement along the catheter between two marks on
the ruler and compare your result with the actual distance from the ruler.
10.5 QCA / QVA
The QCA and QVA applicaons have similar tasks, and they are described together in the following
secons.
Quantave Coronary Analysis (QCA)
You use QCA to mark the contours of a coronary artery in the heart, analyze a stenosis, and create,
store and print reports of the analysis.
Quantave Vascular Analysis (QVA)
You use QVA to mark the contours of aorc and peripheral arteries, analyse a stenosis, and create,
store and print reports of the analysis.
10.5.1 QCA / QVA Tasks
A set of predened tasks is used to ensure coronary or vascular analysis is performed in a logical way.
The QCA and QVA applicaons provide the following tasks in order:
• Select Series
• Calibraon
• Analysis
• Result
When a series is selected, the system progresses automacally to the Calibraon task.
2D Quantave Analysis (Opon) QCA / QVA
Azurion Release 1.2 Ins
trucons for Use 160 Philips Healthcare 4522 203 52421