NOTE Ensure that the ES image that you select is in the same cardiac cycle as the ED image that you
select
ed in the End Diastole task.
1 Click the End Systole task.
2 Use the navigaon toolbar to review the series and select an image that shows the ES cardiac
phase.
Dening the ES Contour Semi-Automacally in LVA
To dene a contour semi-automacally in LVA, you place three key points on the selected image.
Aer dening the ES contour, both the ED and ES contours are displayed in each image in the series.
The contours are highlighted when you view the image used to dene the contour.
Figure 77 L
VA semi-automac ES contour detecon
The main analysis results are displayed in a panel in the lower right corner.
1 Click Semi-Automac in the control panel.
2 Click on the superior border of the aorc root.
3 Click on the inferior border of the aorc root.
4 Click on the apex.
5 If you are using Biplane LVA: Perform this process on both the frontal image and the lateral image
so that the contour is detected in each image.
6 If you want to delete the contour and start over, click Delete in the control panel.
7 Use the navigaon toolbar to check the accuracy of the ED and ES contours in each image of the
series.
2D Quant
ave Analysis (Opon) LVA / RVA
Azurion Release 1.2 Ins
trucons for Use 174 Philips Healthcare 4522 203 52421