• Fir
st image: displays the rst image in the series.
• Middle image: displays the middle image in the series.
The default selecon is Middle image.
4 To change the way that navigaon works when you reach the beginning or end of a study, select a
new seng in the Image Navigaon Model list.
The sengs available are:
• Navigate images in all series: image navigaon does not stop at the end of the current series,
but connues to the next available series for in the selected study.
• Stop at the beginning and at the end of the series: image navigaon stops at the beginning or
end of the current series.
• Step through the images in a loop: image navigaon of the current series connues unl
stopped.
5 To specify a maximum length of me for series image replay, enter a value in seconds for Replay
Time Out.
6 To specify a maximum length of me for study image replay, enter a value in seconds for Study
Cycle Replay Time Out.
7 To change the way that angles are displayed on the system, select the angle avor in the Rotaon/
Angulaon Display Flavor list.
The sengs available are:
• Cardio (LAO/RAO, CRAN/CAUD)
• Vascular (Rot, Ang)
8 To undo any changes you have made, click Undo Changes.
9 To save your changes, click Save.
NOTE The changes take eect aer the next system shutdown and startup.
10 To close the System Customizaon window, click Close.
12.13 Changing Display Preferences
To ensure the correct mouse movements between screens, you can select the control monitor
conguraon you are using.
You can also specify a waing me for screen saver acvaon.
1 On the System menu, click Customizaon to display the System Customizaon window.
2 In the X-Ray Applicaon sengs group, click Viewing and Processing.
3 To ensure the correct mouse movements between the acquision and review windows, select the
conguraon you are using in the Displays and Mouse Control sengs.
4 To change the waing me before the screen saver is acvated, select a suitable me in the Screen
Saver Wait Time list.
5 To immediately acvate the screen saver, click Acvate Screen Saver.
Moving the mouse or pressing any key on the keyboard will deacvate the screen saver.
6 To undo any changes you have made, click Undo Changes.
User Cus
tomizaon Changing Display Preferences
Azurion Release 1.2 Ins
trucons for Use 226 Philips Healthcare 4522 203 52421