EasyManua.ls Logo

Philips Azurion - 16.20.4 Examples of Settings with a Relatively High Air Kerma (Rate)

Philips Azurion
402 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluoroscopy Flavors: Head (ClarityIQ)
Field of View: 0 1 2 3 4 5 6 7
Field Size (cm): 48 42 37 31 27 22 19 15
Flavor Paent T
ype Ref. AK (mGy/s)
Low Default 0.102 0.127 0.136 0.151 0.175 0.205 0.230 0.260
Medium Default 0.155 0.194 0.213 0.244 0.294 0.348 0.388 0.433
Normal Default 0.283 0.350 0.383 0.438 0.501 0.584 0.668 0.799
Roadmap Modes (Vessel Phase): Head
Field of View: 0 1 2 3 4 5 6 7
Field Size (cm): 48 42 37 31 27 22 19 15
Mode Paent T
ype Ref. AK (mGy/image)
Navigate Default 0.202 0.251 0.274 0.311 0.356 0.414 0.443 0.447
Carod Default 0.178 0.212 0.228 0.254 0.285 0.323 0.344 0.347
Coil Default 0.857 1.119 1.254 1.478 1.753 2.129 2.520 3.147
Roadmap Modes (Vessel Phase): Head (ClarityIQ)
Field of View: 0 1 2 3 4 5 6 7
Field Size (cm): 48 42 37 31 27 22 19 15
Mode Paent T
ype Ref. AK (mGy/image)
Navigate Default 0.081 0.098 0.106 0.120 0.135 0.155 0.175 0.207
Coil Default 0.155 0.203 0.227 0.268 0.318 0.385 0.455 0.569
UnSubtract Default 0.081 0.098 0.106 0.120 0.135 0.155 0.175 0.207
16.20.4 Examples of Sengs with a R
elavely High Air Kerma (Rate)
The following table shows examples of exposure procedures that produce a relavely high reference air
kerma value, compared to other procedures, for the dierent Azurion systems (according to IEC
60601-2-54:2009, 203.5.2.4.5.101b 4):
System X-ray Protocol Field size Paent type
C12/F12 Head Subclavian 3 fps 15 cm Default
C20/F20 Head Subclavian 3 fps 15 cm Default
C12/F12 with ClarityIQ (op-
on)
Head Cerebral 2 fps Normal 15 cm Default
C20/F20 with ClarityIQ (op-
on)
Head Cerebral 2 fps Normal 15 cm Default
The following table shows examples of uor
oscopy avors that produce a relavely high Reference Air
Kerma value, compared to other procedures, for the dierent Azurion systems (according to IEC
60601-2-54:2009, 203.5.2.4.5.101b 3):
System X-ray Protocol Flavor Field size Paent type
C12/F12 Head High 15 cm Default
C20/F20 Head High 15 cm Default
C12/F12 with ClarityIQ (op-
on)
Head Normal 15 cm Default
Technical In
formaon Typical Reference Air Kerma (Rate) Values
Azurion Release 1.2 Ins
trucons for Use 310 Philips Healthcare 4522 203 52421

Table of Contents