5 T
o hide or show the contour of the catheter as you work, select or clear Hide in the control panel.
6 To edit a contour, click Edit in the control panel, and do one of the following:
• Click along the walls of the catheter in the image and then double-click on the last posion to
complete the contour.
• Drag along the walls of the catheter in the image to correct the posion of the contour.
7 If you are using Biplane LVA/RVA: Trace the centerline of the catheter in both the frontal image and
the lateral image.
8 When the contours are complete, select the catheter size from the list in the control panel.
If the desired catheter size is not available, you can type it directly in the box.
9 You can delete the calibraon at any me and start over by clicking Delete in the task panel.
10 To accept the calibraon factor, click Accept and Connue in the control panel.
Distance Calibraon
You perform distance calibraon by marking a known distance in the image.
1 Use the navigaon toolbar to review the series and select an image to be used for calibraon.
NOTE You can change the calibraon image at any me by clicking Change in the control panel
and selecng a dierent image.
2 Click the Calibraon task.
3 Click Distance in the Select calibraon method list.
4 Click Draw in the control panel and do the following:
• Click in the image at the desired start point of the line.
• Click again at the desired end point.
5 To hide or show the line, select or clear Hide in the control panel.
6 To edit the line, click Edit in the control panel, and do the following:
a Move the pointer over the start point or the end point
b Drag the point to a new posion.
7 If you are using Biplane LVA/RVA: Mark the line in both the frontal image and the lateral image.
8 You can delete the calibraon at any me and start over by clicking Delete in the task panel.
9 Aer drawing the line, select the distance in the list in the control panel.
If the desired distance is not available, you can type it directly in the box.
10 To accept the calibraon factor, click Accept and Connue in the control panel.
Sphere Calibraon
You perform sphere calibraon by idenfying a sphere of a known size in the image.
2D Quant
ave Analysis (Opon) LVA / RVA
Azurion Release 1.2 Ins
trucons for Use 170 Philips Healthcare 4522 203 52421