Philips Azurion - 5.9.2 Positioning the C-arm

Philips Azurion
402 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If the foot switch is ed with a s
terile cover, do not t the cover too ghtly. This is to ensure that when
one pedal is pressed the cover does not acvate other pedals.
Spillage of Liquids
Prevent spillage of liquids, which may bring live parts of the equipment into contact with conducve
enclosures or direct contact with the operator, other personnel, or the paent.
5.9.2 Posioning the C-arm
You posion the C-arm in the working posion using the control module.
The following working posions are available:
Head side
Doctor side
Nurse side
Rotaon and angulaon movements are motorized with variable speed. The greater the deecon of
the switch on the control module, the faster the C-arm moves. For normal operaon these speeds are 0
- 25 degrees per second (lower maximum speeds can be congured by technical support). If the stand
is not in its working range the maximum speeds are reduced to 8 degrees per second.
The direcon of movement is aected by the posion of the Orientaon switch on the underside of the
control module. For more informaon, see Selecng the Tableside Posion for the Control
Module (page 59).
The following procedure assumes the control module is posioned at the doctor side of the table.
Check the table lock status, and lock or unlock the table as appropriate during the procedure. For more
informaon, see Locking and Unlocking C‐arm and Table Movements (page 90).
Figure 31 P
osioning the C-arm at the head end (le) and nurse side (right)
1 If the table has the lt opon installed, ensure that the table is not lted.
2 If the table has the pivot opon installed, ensure that the table is not pivoted.
3 Set the detector source-to-image distance to the maximum.
4 Place the paent on the table in the desired posion.
For more informaon, see Posioning the Paent on the Table (page 52).
5 Move the C-arm to the desired posion.
Preparing a P
aent Study Preparing the System
Azurion Release 1.2 Ins
trucons for Use 57 Philips Healthcare 4522 203 52421