The oponal K
eep Maximum SID funcon keeps the detector in the maximum SID posion, to prevent
it from hing objects while performing APC movements. Recalling a stand posion may cause a
collision. If necessary, stop the recall funcon and posion the stand manually.
3.1.5 Collision Prevenon Override
You can override collision prevenon funcon in the following situaons.
Smart BodyGuard Override
You can override the BodyGuard funcon if it is blocking motorized stand movements that may be
caused by equipment used around the paent and table, for example ECG cables. This funcon is called
Smart BodyGuard Override. It is the responsibility of the operator to ensure that collisions with the
paent or equipment do not occur while the override funcon is acve.
To acvate the Smart BodyGuard Override funcon aer a movement is blocked by the BodyGuard,
release the movement switch and reacvate it within 5 seconds.
When you acvate the override funcon, a message is displayed in the status area, and a repeang
beep sound can be heard. The maximum movement speed during override is reduced compared to
normal movements. The override funcon is deacvated and normal movements are available again if
the requested movement is no longer being limited by the BodyGuard sensor.
NOTE Smart BodyGuard Override is a congurable funcon and it may not be enabled on every
system
Table Height Collision Override
If you need to perform CPR, you can override the force sensor that stops table height and lt
movements.
The table is equipped with a force sensor measuring the force vercally applied to the surface of the
tabletop. Normally the measured force is determined by the paent weight. During motorized
movement, if a collision force exceeding the safety threshold is detected, the movement stops and is
briey reversed.
To override the force sensor, release the table lt switch or table height switch and then re-apply the
switch movement within 5 seconds to connue the movement. If no switch movements are performed
within 5 seconds, override mode is deacvated and table movements are stopped.
NOTE There are no audible signals during table height collision override mode.
3.1.6 Intelligent Collision Protecon
Intelligent Collision Prevenon (iCP) prevents collisions between the tabletop, the X-ray tube, and the
stand. In a biplane system, iCP also prevents collisions between the frontal stand and the lateral stand.
When the distance between the tabletop and stand becomes too small, a collision is prevented by
stopping motorized stand movements (except for motorized detector shi movement).
The iCP funcon allows you to resume movement and decrease the distance between the table and
stand in a controlled way:
• A small movement step is made when the calculated distance decreases further during the
movement.
• When the calculated distance remains the same during the movement, the movement connues at
reduced speed.
• When the calculated distance increases during the movement, the movement connues at normal
speed.
About the System Equipment in the Examinaon Room
Azurion Release 1.2 Ins
trucons for Use 31 Philips Healthcare 4522 203 52421